PCU-F3000 Pumpless & Tankless Chiller

PCU-F3000 dari Apiste adalah chiller pumpless dan tankless (tanpa pompa internal & tangki), desain fleksibel untuk dipasang ke sistem sirkulasi air yang sudah ada. Mampu mendinginkan 2.800 / 3.200 W (mode 50 / 60 Hz), dengan stabilitas suhu ±0,1 °C. Rentang aliran air sirkulasi dari 10 hingga 60 ℓ/min, dengan fixed flow rate 15 ℓ/min, lingkungan kerja ideal 5-45 °C, kontrol temperatur menggunakan sensor outlet air atau eksternal.

Ukuran unit sekitar H697 × W400 × D600 mm; berat kira-kira 51 kg.

Spesifikasi PCU-F3000 Pumpless & Tankless Chiller

Cooling scheme Air-cooling type
Cooling capacity (W) *1 2800 / 3200
Circulated water *2 Clean water, Ethylene glycol solution (concentration 15 wt% or less), Propylene glycol solution (concentration 15 wt% or less)
Temperature stability (℃) *3 ±0.1
Use Area Temperature (℃) 5 to 45
Temperature control / Fixed system Control target *4 : Outlet liquid temperature or External temperature sensor measurement value
Range[℃] : 5 to 40
Temperature control / Synchronized system Reference *4 : External temperature sensor measurement value
Control target : Outlet liquid temperature
Synchronization range [℃] : -10.0 to +10.0 relative to the reference temperature (factory default setting: 0.0)
Circulated water flow rate range (L/min) 10 to 60
Circulating water fixed flow rate (ℓ/min) 15/15
Circulating water maximum operating pressure (MPa) 1
Dimensions (mm) *5 H697×W400×D600
Coating Color Munsell N-87
Weight (kg) Approx. 51
Casters (Φmm×unit) 50×4
Rated Voltage (V) *6 3-phase 200 VAC
Power Consumption (kW) *1 0.7 / 1.0
Current consumption (A) *1 2.7 / 3.3
Recommended breaker capacity (A) 10
Circulating water inlet / outlet port size Rc3/4
Drain discharge port size G1/2
Components / Compressor Fully sealed rotary type
Components / Condenser Parallel flow condenser
Components / Evaporator Structure : Plate-type heat exchanger
Material : SUS316 (copper brazing)
Refrigerator R1234yf 340g
Noise (dB) *7 51 / 54
Safety circuit / Compressor Thermal protector, electromagnetic switch
Safety circuit / Fan motor Thermal protector, current fuse
Input / Alarm output No voltage, normally open (maximum 250V AC and 1 A/30 V DC and 2A)
Input/Output / Water temperature upper / lower limit output No voltage, normally open (maximum 250V AC and 1 A/30 V DC and 2A)
Input/Output / Operation stop output No voltage, normally open (maximum 250V AC and 1 A/30 V DC and 2A)
Input/Output / Emergency stop input No-voltage input
Input/Output / Operation stop input No-voltage input
Input/Output / Error reset input No-voltage input
Input/Output / External temperature sensor *7 Pt100